T3PM1100 User Manual
124
SCPI Commands
*CLS ......................................................................... 124
*ESE ......................................................................... 124
*ESR ......................................................................... 125
*IDN ........................................................................ 125
*OPC ........................................................................ 125
*OPT ........................................................................ 126
*RST ......................................................................... 126
*SRE ......................................................................... 126
*STB ......................................................................... 127
*TRG ........................................................................ 127
*CLS
Description
Clears the standard event status register, extended
event status register, and error queue.
Syntax
*CLS
*ESE
Description
Sets or returns the ESER (Event Status Enable
Register) contents.
Syntax
Query Syntax
*ESE <NR1>
*ESE?
Parameter/
Return parameter
<NR1>
0~255
Example
*ESE 65
Set the ESER to 01000001
*ESE?
->130
ESER=10000010
Set
Set
Query
Содержание Everywhereyoulook T3PM1100
Страница 1: ...T3PM1100 Digital Power Meter User Manual ...
Страница 75: ...MEASUREMENT AND OTHER FUNCTIONS 75 3 Press ESC button to return back to original display mode ESC ...
Страница 119: ...COMMAND OVERVIEW 119 CR LF The most common EOL character is CR LF Message Separator EOL or semicolon Command Separator ...
Страница 191: ...COMMAND OVERVIEW 191 Example STATUS QMESSAGE ON STATUS QMESSAGE STATUS QMESSAGE 1 ...
Страница 209: ...APPENDIX 209 Status system The diagram below is a description of the status system ...
Страница 213: ...APPENDIX 213 Dimensions Units mm ...