T3PM1100 User Manual
200
Input
Item
Specifications
Input type
Voltage Floating input through resistive voltage
divider
Current
Floating input through shunt
Measure range
Voltage
15 V, 30 V, 60 V,150 V, 300 V, 600 V
Current
Direct
input
5 mA, 10 mA, 20 mA, 50 mA, 100 mA,
200 mA, 0.5 A, 1 A, 2 A, 5 A, 10 A, 20 A
Sensor
input
EX1: 2.5 V, 5 V, 10 V
EX2: 50 mV, 100 mV, 200 mV, 500 mV, 1
V, 2 V
Input impedance
Voltage
Input resistance: approach 2 M
Ω
Current
Direct input range 5 mA
~ 200 mA
Input resistance:
approach 505 m
Ω
Direct input range 0.5A
~ 20 A
Input resistance:
approach 5 m
Ω
Sensor input
Input range 2.5 V ~ 10 V
(EX1)
Input resistance:
approach 100 k
Ω
Input range 50 mV ~ 2 V
(EX2)
Input resistance:
approach 20 k
Ω
Continuous maximum
allowable input
Voltage
peak value of 1.5 kV or RMS value of 1
kV, whichever is less
Current
Direct input range 5 mA
~ 200 mA
peak value of 30 A or
RMS value of 20 A,
whichever is less
Direct input range 0.5A
~ 20 A
peak value of 100 A or
RMS value of 30 A,
whichever is less
Sensor
input
peak value less than or equal to 5 times
of the rated range
Input bandwidth
DC, 0.1 Hz ~ 100kHz
Continuous maximum
Common-mode voltage
600 Vrms, CAT
Ⅱ
Line filter
select OFF or ON (cut off frequency of 500 Hz)
Frequency filter
select OFF or ON (cut off frequency of 500 Hz)
A/D converter
Simultaneous conversion voltage and current inputs
Resolution
16bits
Maximum conversion rate
Approx. 300kHz
Содержание Everywhereyoulook T3PM1100
Страница 1: ...T3PM1100 Digital Power Meter User Manual ...
Страница 75: ...MEASUREMENT AND OTHER FUNCTIONS 75 3 Press ESC button to return back to original display mode ESC ...
Страница 119: ...COMMAND OVERVIEW 119 CR LF The most common EOL character is CR LF Message Separator EOL or semicolon Command Separator ...
Страница 191: ...COMMAND OVERVIEW 191 Example STATUS QMESSAGE ON STATUS QMESSAGE STATUS QMESSAGE 1 ...
Страница 209: ...APPENDIX 209 Status system The diagram below is a description of the status system ...
Страница 213: ...APPENDIX 213 Dimensions Units mm ...