BLZZRD Portable Refrigerated Sampler
Section 4 Standard Programming
4-14
Example: Programming with 750 Area Velocity Module Installed
1. With the sampler off, insert the module in the bay
on the controller.
2. Turn the sampler on by pressing the On/Off key.
Press
. Download stored data, if you want to keep
it. Pressing DONE erases it and advances to the
next screen.
3. The startup screen appears for 8 seconds.
4. The option PROGRAM will be blinking. Press
.
5. The option NO will be blinking. Press
.
6. Select the desired temperature units.
7. Select the desired units for the flow rate. For this
example, choose cfs. Press
.
8. Select the desired units for the flow volume. For
this example, choose Mgal. Press
.
9. Select the desired units for the flow rate. For this
example, choose fps. Press
.
10. Select YES to program the module.
11. The 750 has two modes of operation:
FLOWMETER or LEVEL AND VELOCITY. For this
example, choose FLOWMETER.
12. When you select Flowmeter, you will have two
options for performing the flow conversion: AREA *
VELOCITY or LEVEL TO FLOW RATE. For this
example, choose AREA * VELOCITY.
13. For this example, choose Channel Shape.
14. For the channel shape, select Round Pipe.
15. Enter the diameter of the round pipe. This example
shows a two-foot diameter pipe.
Continued…
MODULE INSERTED---
DOWNLOAD DATA NOW
OR LOSE ALL DATA!
DONE
BLZZRD
STANDARD PROGRAMMING
For HELP at any
screen press ? key.
RUN
PROGRAM
VIEW REPORT
OTHER FUNCTIONS
SITE DESCRIPTION:
“FACTORY051“
CHANGE?
YES
NO
SELECT UNITS FOR
TEMPERATURE
°F
°C
SELECT UNITS FOR
FLOW RATE:
cfs
gps gpm Mgd
lps m3s m3h m3d
SELECT UNITS FOR
FLOW VOLUME:
cf gal
Mgal
m3 lit
SELECT UNITS FOR
VELOCITY:
fps
mps
PROGRAM MODULE?
YES
NO
MODE OF OPERATION:
FLOWMETER
LEVEL AND VELOCITY
FLOW CALCULATION:
AREA*VELOCITY
LEVEL TO FLOW RATE
LEVEL TO AREA:
CHANNEL SHAPE
DATA POINTS
ROUND PIPE
U-CHANNEL
RECTANGULAR CHANNEL
TRAPEZOIDAL CHANNEL
ROUND PIPE
DIAMETER =
2.00
ft
Содержание Everywhereyoulook BLZZRD
Страница 2: ......
Страница 4: ......
Страница 8: ...BLZZRD Portable Refrigerated Sampler Safety vi ...
Страница 16: ...BLZZRD Portable Refrigerated Sampler Table of Contents xiv ...
Страница 34: ...BLZZRD Portable Refrigerated Sampler Section 1 Introduction 1 18 ...
Страница 58: ...BLZZRD Portable Refrigerated Sampler Section 2 Installation Preparation 2 24 ...
Страница 72: ...BLZZRD Portable Refrigerated Sampler Section 3 Programming Introduction 3 14 ...
Страница 88: ...BLZZRD Portable Refrigerated Sampler Section 4 Standard Programming 4 16 ...
Страница 219: ...BLZZRD Portable Refrigerated Sampler Appendix B Material Safety Data Sheets B 8 ...
Страница 233: ...BLZZRD Portable Refrigerated Sampler Appendix D Replacement Parts D 6 ...
Страница 241: ...BLZZRD Portable Refrigerated Sampler Appendix F Battery Selection Guide F 4 ...