BLZZRD Portable Refrigerated Sampler
Section 5 Extended Programming
5-29
If the Liquid Detector is off:
•
You must enter the SUCTION HEAD.
•
No RINSES or RETRIES will be available.
•
The COUNT TO LIQUID column of the RESULTS
report will read “0.”
•
A complete purge will occur between samples when
using multiple bottles-per-sample distribution.
5.15.2 Programming Style
The sampler has two different styles of programming screens:
normal and quick view. These are both explained in Section 3.
When you are in extended programming, the default style is
Quick View, which allows you to move more quickly through the
screens. If you prefer a step-by-step display for the screens,
switch to the Normal setting.
5.15.3 Measurement Interval
If using a 700 Series module or SDI-12 Sonde parameter for
enabling, the sampler will request a Measurement Interval.
The Measurement Interval allows the sampler to compare the
parameter to the enabling conditions at a rate faster than the
Data Storage interval. If the Measurement Interval is equal to or
greater than the Data Storage interval, the parameter will be
checked at the rate of Data Storage interval.
Normally, there are two considerations when selecting the Mea-
surement Interval: power consumption and duration of the
enabling event. This setting for battery powered samplers will
require some degree of balance between the need to conserve
power yet detect the enabling event before it has passed.
5.15.4 Dual Sampler Mode
Dual Sampler Mode lets you operate two samplers as a pair. The
primary sampler disables the secondary one until the primary
sampler completes its sampling program. The samplers then
reverse roles with the second sampler becoming the primary one.
Setting Up Dual Samplers
–
1. Turn on the dual sampler mode in both samplers.
2. Install the samplers.
3. Connect the samplers with the Dual Sampler Mode Inter-
connect Cable by attaching the cable to the sampler’s
flow-meter connectors. If your installation includes a flow
meter or flow logger, use the Dual Sampler Mode to Flow
Meter Cable instead.
4. Decide which sampler is to be the primary sampler; the
primary sampler should be the sampler with the program
you want to run first.
5. Turn the primary sampler on and run its sampling pro-
gram by selecting RUN. After at least 5 seconds, turn the
secondary sampler on and select RUN from its main menu.
USE LIQUID DETECTOR?
YES
NO
MEASUREMENT INTERVAL
5
15
30
SECONDS
1
2
5
MINUTES
STORAGE INTERVAL
DUAL SAMPLER MODE?
YES
NO
Содержание Everywhereyoulook BLZZRD
Страница 2: ......
Страница 4: ......
Страница 8: ...BLZZRD Portable Refrigerated Sampler Safety vi ...
Страница 16: ...BLZZRD Portable Refrigerated Sampler Table of Contents xiv ...
Страница 34: ...BLZZRD Portable Refrigerated Sampler Section 1 Introduction 1 18 ...
Страница 58: ...BLZZRD Portable Refrigerated Sampler Section 2 Installation Preparation 2 24 ...
Страница 72: ...BLZZRD Portable Refrigerated Sampler Section 3 Programming Introduction 3 14 ...
Страница 88: ...BLZZRD Portable Refrigerated Sampler Section 4 Standard Programming 4 16 ...
Страница 219: ...BLZZRD Portable Refrigerated Sampler Appendix B Material Safety Data Sheets B 8 ...
Страница 233: ...BLZZRD Portable Refrigerated Sampler Appendix D Replacement Parts D 6 ...
Страница 241: ...BLZZRD Portable Refrigerated Sampler Appendix F Battery Selection Guide F 4 ...