BLZZRD
™
Portable Refrigerated Sampler
Section 2 Installation/Preparation
2-21
Flow Meter Connector
–
Connect the sampler to a Teledyne
ISCO 4200 Series Flow Meter, 4100 Series Flow Logger, 2100
Series Flow Module, 1640 Liquid Level Actuator, or input
interface by attaching their connect cable to the Flow Meter Con-
nector. When using the sampler with an external instrument,
remember these guidelines:
•
Both the sampler and the flow instrument must be on
and running a program.
•
If using a flow meter, logger, or module for flow pacing,
the sampler and the flow instrument must both run
programs with flow pacing settings.
•
If using a flow meter or flow logger for trigger pacing,
the sampler must run a program with flow pacing
settings, and the flow instrument must run a program
with trigger pacing settings.
•
When the sampler runs an event-paced program, it
disregards pacing pulses from a flow meter or flow
logger. However, the sampler continues to monitor for
enable signals.
WARNING
Most Teledyne ISCO flow meters are not certified for use in
hazardous locations as defined by the National Electrical
Code. Contact your Teledyne ISCO representative for your
equipment’s status. Never operate equipment in hazardous
locations that has not been certified as safe for such use.
Refer to Appendix C in the back of this manual for more
safety information.
2.10 Programming the
BLZZRD
Due to the many features supported by the BLZZRD sampler,
programming instructions are not addressed in this section.
Refer to the following sections for programming instructions:
•
Section 3 introduces the keypad and display and
provides general programming instructions.
•
Section 4
provides instructions for the sampler’s
Standard
programming features.
•
Section 5
provides instructions for the sampler’s
Extended
programming features.
2.11 Locking the Sampler
The BLZZRD sampler contains two locking latches by which the
sampler may be mechanically locked. Use padlocks or a locking
cable mechanism to tamper-proof the sampler.
2.11.1 Program Lock
A program lock software option can be purchased from Teledyne
ISCO. This option secures access to the sampler’s program set
-
tings. The program lock is ordered separately as Teledyne ISCO
Part Number 60-9003-532.
Содержание Everywhereyoulook BLZZRD
Страница 2: ......
Страница 4: ......
Страница 8: ...BLZZRD Portable Refrigerated Sampler Safety vi ...
Страница 16: ...BLZZRD Portable Refrigerated Sampler Table of Contents xiv ...
Страница 34: ...BLZZRD Portable Refrigerated Sampler Section 1 Introduction 1 18 ...
Страница 58: ...BLZZRD Portable Refrigerated Sampler Section 2 Installation Preparation 2 24 ...
Страница 72: ...BLZZRD Portable Refrigerated Sampler Section 3 Programming Introduction 3 14 ...
Страница 88: ...BLZZRD Portable Refrigerated Sampler Section 4 Standard Programming 4 16 ...
Страница 219: ...BLZZRD Portable Refrigerated Sampler Appendix B Material Safety Data Sheets B 8 ...
Страница 233: ...BLZZRD Portable Refrigerated Sampler Appendix D Replacement Parts D 6 ...
Страница 241: ...BLZZRD Portable Refrigerated Sampler Appendix F Battery Selection Guide F 4 ...