Table 8-4 BLZZRD Remote Phone Commands
Phone
Command
Name
Description
Start
This command will immediately start any program set to WAIT FOR PHONE
CALL and currently waiting for the remote start command (see Section 4.6).
Load and Run
Program 1
This command will load Program 1 as the sampler’s current Extended pro
-
gram, reset partition if the storage interval has been changed, and run the pro-
gram. RUN1 is valid when the sampler is in standby, or when it is waiting for a
start command. RUN1 will also run the current program while in Standard pro-
gramming.
Load and Run
Program 2, 3, or 4
These commands will load the selected program number as the sampler’s cur
-
rent Extended program, reset partition data if the storage interval has
changed, and run the program. They are valid when the sampler is in standby,
or when it is waiting for a start command. The commands are available for
Extended programming only.
Disable
This command is similar in functionality to a pin F low signal (a signal typically
used to enable or disable the sampler from a flow meter). At run time, the sam-
pler starts with the phone enable in the enabled state. PHONE DISABLE is
valid while a program is running. If successful, a reply of REMOTE ENABLE IS
FALSE will be returned.
A Remote Disable is the highest priority disable. If the sample is disabled with
this command, no condition other than command 6#, ENABLE, will allow the
program to become enabled again.
Enable
This command is used to re-enable a phone-disabled sampler. PHONE
ENABLE is valid while a program is running. If successful, a reply of REMOTE
ENABLE IS TRUE will be returned.
Grab Sample
This command causes the sampler currently running a program to take a sam-
ple. The sample will count as one of the program’s samples and is placed in
the current bottle (Part A’s current bottle when using two-part programming).
TAKE_SAMPLE is valid while the program is running, after the start time, but
not while in the MANUAL PAUSE screens. If successful, a response similar to
PUMPING XXML SAMPLE INTO BOTTLE ## will be returned. If not, a
response similar to NO SAMPLE, PROGRAM WAITING START will be
returned.
0 #
1 #
2 #
3 #
4 #
5 #
6 #
7 #
Содержание Everywhereyoulook BLZZRD
Страница 2: ......
Страница 4: ......
Страница 8: ...BLZZRD Portable Refrigerated Sampler Safety vi ...
Страница 16: ...BLZZRD Portable Refrigerated Sampler Table of Contents xiv ...
Страница 34: ...BLZZRD Portable Refrigerated Sampler Section 1 Introduction 1 18 ...
Страница 58: ...BLZZRD Portable Refrigerated Sampler Section 2 Installation Preparation 2 24 ...
Страница 72: ...BLZZRD Portable Refrigerated Sampler Section 3 Programming Introduction 3 14 ...
Страница 88: ...BLZZRD Portable Refrigerated Sampler Section 4 Standard Programming 4 16 ...
Страница 219: ...BLZZRD Portable Refrigerated Sampler Appendix B Material Safety Data Sheets B 8 ...
Страница 233: ...BLZZRD Portable Refrigerated Sampler Appendix D Replacement Parts D 6 ...
Страница 241: ...BLZZRD Portable Refrigerated Sampler Appendix F Battery Selection Guide F 4 ...