BLZZRD Portable Refrigerated Sampler
Section 5 Extended Programming
5-28
Note that if you have recently calibrated a parameter probe, the
sampler may delay the start and display “PLEASE WAIT” until
valid readings are expected from the probe. The delay accounts
for the time typically required to install the probe, and lasts for
five minutes after you complete the calibration.
5.13.1 Run Time Screens
Run time screens, which report the program’s status while
running a program, are explained in Section 6.
5.13.2 Interrupting a
Running Program
Interrupting a running program is also explained in Section 6.
The scrolling menu that appears during the manual paused state
can have two extra options when using extended programming:
Adjust Enable
–
This option is available for extended sampling
programs that have programmed enable conditions. Selecting
this option allows you to change the set-points or ranges for the
enable conditions. The sampler displays all of the programming
screens that define the thresholds of the programmed enables.
You cannot change the type of enables used or the way they are
combined.
If rain is an enabling condition, you also have the option of
resetting the rain history.
Adjust Suction Head
–
If you have chosen to manually enter
the suction head, and are in the manual paused state, the
scrolling menu will have an option to adjust the suction head. If
you make changes to the suction head value, the sampler logs a
changed suction head event.
5.14 Other Functions
In extended programming, the “Other Functions” menu screen
includes options for:
•
Maintenance (see Section 9)
•
Manual Functions (see Section 6)
•
Software Options (see Section 5.15)
•
Hardware (see Section 5.16)
5.15 Software Options
The software option screens allow you to set up a number of dif-
ferent software options:
5.15.1 Liquid Detector
Enable/Disable
Under some conditions (pressurized lines, sampling foam), a
more repeatable sample volume may be delivered by disabling
the liquid detector. Selecting NO will turn the Liquid Detector
off.
•
Liquid Detector Enable
•
Display Backlighting
•
Full Bottle Detection
•
Event Marks
•
Measurement Interval
•
Pump Purge Counts
•
Dual Sampler Mode
•
Serial Output
•
Normal/Quick View
Screens
•
Interrogator Connect
Power
Содержание Everywhereyoulook BLZZRD
Страница 2: ......
Страница 4: ......
Страница 8: ...BLZZRD Portable Refrigerated Sampler Safety vi ...
Страница 16: ...BLZZRD Portable Refrigerated Sampler Table of Contents xiv ...
Страница 34: ...BLZZRD Portable Refrigerated Sampler Section 1 Introduction 1 18 ...
Страница 58: ...BLZZRD Portable Refrigerated Sampler Section 2 Installation Preparation 2 24 ...
Страница 72: ...BLZZRD Portable Refrigerated Sampler Section 3 Programming Introduction 3 14 ...
Страница 88: ...BLZZRD Portable Refrigerated Sampler Section 4 Standard Programming 4 16 ...
Страница 219: ...BLZZRD Portable Refrigerated Sampler Appendix B Material Safety Data Sheets B 8 ...
Страница 233: ...BLZZRD Portable Refrigerated Sampler Appendix D Replacement Parts D 6 ...
Страница 241: ...BLZZRD Portable Refrigerated Sampler Appendix F Battery Selection Guide F 4 ...