BLZZRD Portable Refrigerated Sampler
Section 7 SDI-12 Sondes
7-9
•
DO - For dissolved oxygen, you can choose between two
methods: DO Standard and Saturation. The saturation
method requires a temperature reading between 0.0°
and 47.0° C. You will also need to enter your altitude
and barometric pressure.
•
ORP - You will asked to enter the value of your
calibration standard.
When calibration is complete, select DONE to proceed to the val-
idation screen. If you do not want to validate any parameters,
select DONE again to return to the Hardware option screens.
7.7.2 Validation Screen
The validation screen appears after you select DONE on the cali-
bration screen. It will be similar to the screen shown below.
Select the desired parameter to validate. At the prompt, place
the probe in the standard. When the reading is stable, press
.
The reading will appear at the bottom of the screen.
Select DONE to return to the Hardware option screens.
7.8 Disconnecting Sondes
When you have disconnected all sondes, and have interrogated
the sampler for all its data, you can indicate that there is no
SDI-12 sonde.
From the main menu, select Other Functions, Hardware.
Advance through the screens until you see one that displays
SDI-12 DATA. Select this screen to display the following:
Select NO SDI-12 SONDE. You will advised that you have a new
hardware setup:
If you have collected all the data, select DONE. A partition
memory message will be displayed briefly, and you will be
returned to the Hardware options screens.
If you do not have a sonde attached, you may want to turn off
automatic scanning for sondes. This will eliminate the time
taken to scan for sondes whenever the sampler is turned on.
SELECT PARAMETER
TO VALIDATE:
SAL5
05DATA5 DO5
ORP5
LEVEL5
DONE
NO SDI-12 SONDE
SETUP KNOWN SONDES
SCAN SDI-12 SONDES
NEW HARDWARE SETUP--
DOWNLOAD DATA NOW
OR LOSE ALL DATA!
DONE
Содержание Everywhereyoulook BLZZRD
Страница 2: ......
Страница 4: ......
Страница 8: ...BLZZRD Portable Refrigerated Sampler Safety vi ...
Страница 16: ...BLZZRD Portable Refrigerated Sampler Table of Contents xiv ...
Страница 34: ...BLZZRD Portable Refrigerated Sampler Section 1 Introduction 1 18 ...
Страница 58: ...BLZZRD Portable Refrigerated Sampler Section 2 Installation Preparation 2 24 ...
Страница 72: ...BLZZRD Portable Refrigerated Sampler Section 3 Programming Introduction 3 14 ...
Страница 88: ...BLZZRD Portable Refrigerated Sampler Section 4 Standard Programming 4 16 ...
Страница 219: ...BLZZRD Portable Refrigerated Sampler Appendix B Material Safety Data Sheets B 8 ...
Страница 233: ...BLZZRD Portable Refrigerated Sampler Appendix D Replacement Parts D 6 ...
Страница 241: ...BLZZRD Portable Refrigerated Sampler Appendix F Battery Selection Guide F 4 ...