BLZZRD
™
Portable Refrigerated Sampler
Section 9 Maintenance
9-6
The example
Resetting the Internal Battery Countdown Timer
shows how to view the expiration date and how to set a new date
after replacing the internal battery. Refer to section 9.7 for
battery replacement instructions.
9.3.4 Diagnostics
The sampler has several self-diagnostic tests that check the
memory (both the RAM and ROM), pump, and the distributor
arm position. It also contains screens that let you reset the
sampler’s program settings and memory to factory settings.
You do not need to run the diagnostic tests on a routine basis.
The diagnostic tests are a troubleshooting tool; Teledyne ISCO’s
Technical Service department staff may ask you to run these
tests when they work with you in diagnosing problems with your
sampler.
Access the diagnostic screens by selecting Other Functions from
the main menu, then Maintenance, and Diagnostics. The unit
will automatically test the memory and display; you can choose
whether it tests the pump and distributor, and whether it reini-
tializes the unit to its original factory settings.
Example: Self-diagnostics
1. From the Main Menu select OTHER FUNCTIONS.
2. Select MAINTENANCE.
3. Select DIAGNOSTICS
4. The controller starts the diagnostics routine by first
testing the RAM (Random Access Memory). If the
controller passes a memory test, it will advance to
the next test. Otherwise it will report a failure and
advise you to contact Teledyne ISCO.
5. The controller continues by testing the ROM
(Read-only Memory).
6. The controller then fills the display, first with all
pixels “on,” then with characters.
If the display is faulty (for example some pixels
always on or always off), the display or its driver
circuitry may need repair. Call Teledyne ISCO. The
diagnostic routine automatically advances to the
next test.
7. Select YES to test the pump or NO to skip this test.
Continued…
RUN
PROGRAM
VIEW REPORT
OTHER FUNCTIONS
MAINTENANCE
MANUAL FUNCTIONS
PROGRAMMING STYLE
SET CLOCK
PUMP TUBE ALARM
INTERNAL BATTERY
DIAGNOSTICS
PLEASE WAIT!
. . . ‘RAM’ PASSED TEST
PLEASE WAIT!
. . . ‘ROM’ PASSED TEST
ABCDEFGHIJKLMNOPQRST
ABCDEFGHIJKLMNOPQRST
ABCDEFGHIJKLMNOPQRST
ABCDEFGHIJKLMNOPQRST
TEST PUMP?
YES
NO
Содержание Everywhereyoulook BLZZRD
Страница 2: ......
Страница 4: ......
Страница 8: ...BLZZRD Portable Refrigerated Sampler Safety vi ...
Страница 16: ...BLZZRD Portable Refrigerated Sampler Table of Contents xiv ...
Страница 34: ...BLZZRD Portable Refrigerated Sampler Section 1 Introduction 1 18 ...
Страница 58: ...BLZZRD Portable Refrigerated Sampler Section 2 Installation Preparation 2 24 ...
Страница 72: ...BLZZRD Portable Refrigerated Sampler Section 3 Programming Introduction 3 14 ...
Страница 88: ...BLZZRD Portable Refrigerated Sampler Section 4 Standard Programming 4 16 ...
Страница 219: ...BLZZRD Portable Refrigerated Sampler Appendix B Material Safety Data Sheets B 8 ...
Страница 233: ...BLZZRD Portable Refrigerated Sampler Appendix D Replacement Parts D 6 ...
Страница 241: ...BLZZRD Portable Refrigerated Sampler Appendix F Battery Selection Guide F 4 ...