BLZZRD Portable Refrigerated Sampler
Section 5 Extended Programming
5-15
Example: Nonuniform Time Pacing
Nonuniform Clock Time Pacing
allows you specify
the exact time at which each sample will be taken.
1. Select CLOCK TIMES.
2. Enter as many as 99 clock times using a 24-hour
format.
Note:
Intervals between samples cannot
exceed 24 hours.
Random Time Pacing
collects samples at random
time intervals.
1. Select RANDOM INTERVALS.
2. Enter the program run time. The controller derives
the number of sample events required for the
program from the distribution settings.
Nonuniform Time Interval Pacing
allows you specify
the time interval between each sample.
1. Select INTERVALS IN MINUTES.
2.
The controller briefly displays…
…then displays the screen to enter as many as 99
nonuniform time intervals. Enter the number of
samples to be taken at each interval, then specify
the interval in minutes.
NONUNIFORM TIME:
CLOCK TIMES
INTERVALS IN MINUTES
RANDOM INTERVALS
TAKE SAMPLES AT:
1. START TIME
2. HH:MM
3. HH:MM
NONUNIFORM TIME:
CLOCK TIMES
INTERVALS IN MINUTES
RANDOM INTERVALS
PROGRAM RUN TIME:
__ HOURS, __ MINUTES
NONUNIFORM TIME:
CLOCK TIMES
INTERVALS IN MINUTES
RANDOM INTERVALS
FIRST SAMPLE AT
START TIME,
THEN . . .
QUANTITY AT INTERVAL
1. __ AT ___ MIN
2. __ AT ___ MIN
3. __ AT ___ MIN
Содержание Everywhereyoulook BLZZRD
Страница 2: ......
Страница 4: ......
Страница 8: ...BLZZRD Portable Refrigerated Sampler Safety vi ...
Страница 16: ...BLZZRD Portable Refrigerated Sampler Table of Contents xiv ...
Страница 34: ...BLZZRD Portable Refrigerated Sampler Section 1 Introduction 1 18 ...
Страница 58: ...BLZZRD Portable Refrigerated Sampler Section 2 Installation Preparation 2 24 ...
Страница 72: ...BLZZRD Portable Refrigerated Sampler Section 3 Programming Introduction 3 14 ...
Страница 88: ...BLZZRD Portable Refrigerated Sampler Section 4 Standard Programming 4 16 ...
Страница 219: ...BLZZRD Portable Refrigerated Sampler Appendix B Material Safety Data Sheets B 8 ...
Страница 233: ...BLZZRD Portable Refrigerated Sampler Appendix D Replacement Parts D 6 ...
Страница 241: ...BLZZRD Portable Refrigerated Sampler Appendix F Battery Selection Guide F 4 ...