Calibrate A/B: This button runs the magazine calibration
procedure. The magazines are calibrated before leaving the
factory and should not need to be calibrated again. However, if
they do need to be recalibrated, perform the following steps
from the Calibrate Magazines screen:
1.
Calibrate one magazine at a time.
2.
With middle vessels removed, place the calibration
block across the center of the middle vessel
brackets, directly beneath the magazine drive screw.
3.
In the Calibrate Magazines screen, touch the
Calibrate button. The magazine will move down and
touch the block. The magazine is now calibrated.
4.
Repeat the above steps for the other magazine.