![Teledyne CD14 Скачать руководство пользователя страница 58](http://html.mh-extra.com/html/teledyne/cd14/cd14_user-manual_1080810058.webp)
Section F – Operation
CD14 Tester & CD AutoPlus Autosampler User Guide—75-108-800 Rev. B—31 May 2021
57 of 130
EAR99 technology subject to restrictions on cover page.
Speed Tolerance: Drive speed tolerance when a method is not
running. The default value is ± 4% of the set point. The range
is 0 (off), 1 – 10%.
Language: Primary user interface language. Currently English
is the only language available.
Printers tab
The Printers tab allows the user to configure the printer(s) the
CD14 Tester instrument will use.
Network Printer: Selected when a network printer is connected
via a TCP/IP network and supports PostScript.
NOTE: Network printers must support PostScript in their
firmware. Printers that support PostScript through PC software
drivers or do not support PostScript will not work. Only the
Epson serial printer provided by Teledyne Hanson is
supported. Both network and serial printers may be enabled at
the same time.
Page Format: Toggles the paper size between letter and A4
when using a PostScript-capable network printer.
NOTE: Setting the wrong paper size may omit lines from the
printout.
I.P. Address: For use when printing to a PostScript-capable
network printer. Allows the user to enter the IP address of the
target printer.
Printer Port: For use when printing to PostScript-capable
network printers. Allows the user to enter the appropriate port
of the target printer (0-65535). The default is 9100.