Section H – CD14 Tester Maintenance
CD14 Tester & CD AutoPlus Autosampler User Guide—75-108-800 Rev. B—31 May 2021
99 of 130
EAR99 technology subject to restrictions on cover page.
Draining the Waterbath
Warning:
Do not set a temperature during this process. Doing
so may result in damage to the heater.
Insert the drain hose in the drain connector on the front left of
the bath. The bath will completely drain approximately 42 L (11
gal).
Guidelines for Cleaning the Waterbath
It is recommended that the bath be cleaned in place.
With a little water in the bath use a soft cloth and wipe the
inside of the bath. Rinse the bath with water and let it drain to
remove foreign material. The better the bath is cleaned the
longer it will stay clean.
Do not use coarse or abrasive materials. They may scratch the
bath. Alcohol based products can cause the bath to crack.
Warning:
The use of chemicals may damage the waterbath
and will not be covered by the warranty.
Adjustments and Calibration
Spindle Shaft Height Adjustment
1.
Install the paddle into the spindle shaft.
2.
Lower the drive head and place a height set gauge
in the bottom of the vessel.
3.
Using a hex key, loosen the bolt on the spindle shaft
clamp.
4.
Gently adjust the shaft up or down so it rests on top
of the height gauge.
5.
Tighten the spindle clamp using the same hex key.
Heater and Temperature Probe Adjustment
1.
Place an NIST traceable thermometer next to the
temperature probe, or in the exit port on the left side
of the bath if adjusting the bath temperature probe.
2.
From the main screen, touch the Tools button on the
main toolbar.