Section F – Operation
CD14 Tester & CD AutoPlus Autosampler User Guide—75-108-800 Rev. B—31 May 2021
72 of 130
EAR99 technology subject to restrictions on cover page.
Time/Alarms
The Time/Alarms button on the main
menu provides access to the instrument
notification alarms and real clock time and date.
Preheat Wakeup tab
This tab allows the user to enable a preheat wakeup alarm.
When triggered, the waterbath heats to the bath temperature
set point at the specified time and date. This allows the tester
to be ready for the start of a workday while saving power
overnight.
Bath Temperature: This is the temperature that the tester will
heat up to. The button brings up numeric keypad. The range is
25 – 51 °C; preset values are 32, 37, and 51 °C.
Alarm Date/Time: The date/time the alarm will begin heating
the bath.
NOTE: The bath temperature is only changed if a test is NOT
currently in progress.
Recurrence: The method by which the alarm repeats. Either
never, at specific days of the week, or at day intervals.