Section K – Specifications
CD14 Tester & CD AutoPlus Autosampler User Guide—75-108-800 Rev. B—31 May 2021
122 of 130
EAR99 technology subject to restrictions on cover page.
CD AutoPlus/AutoFill General Specifications
Sampler Volume Accuracy:
0.5 to 5.0 mL: ± 0.05
5.1 to 10.0 mL: ± 0.10
10.1 to 15.0 mL: ± 0.15
Sampling:
Minimum sample interval: 1 min (5 min is normal)
Maximum sample interval: 999 hours, 59 min (999:59)
Sample time points: 50 maximum
Maximum sample volume: 90% of vial capacity
Collection:
13 x 100 rack: 18 x 8 test tubes, 144 total
16 x 100 rack: 18 x 8 test tubes, 144 total
12 x 32 rack: 21 x 8 vials, 168 total
UPLC rack (12 x 32): 3 trays 6 x 8 vials, 144 total
Septum: Pre-cut type only.
Warning:
Failure to use pre-cut septa can result in damage to the AutoFill needles.