Section C – Regulatory and Safety
CD14 Tester & CD AutoPlus Autosampler User Guide—75-108-800 Rev. B—31 May 2021
18 of 130
EAR99 technology subject to restrictions on cover page.
CD AutoPlus General Safety Considerations
The installation category (overvoltage category) for this
instrument is Level II. The Level II category pertains to
equipment that receives its electrical power from a local level,
such as an electrical wall outlet.
This instrument must be connected to a grounded electrical
outlet. If condensation forms on or inside the AutoFill collector,
allow it to dry thoroughly before connecting to a power source.
Never work on the electrical components in the system while
there is power to the unit.
Warning:
Disconnect power before servicing the instrument.
Review all safety and environmental precautions pertaining to
any chemicals that are to be used in conjunction with this
equipment.
CD AutoPlus CSA Safety Considerations
For indoor use only
Maximum altitude up to 2,000 meters
Environmental operating temperature 5 to 40 °C
Operating relative humidity 80% for temperatures up
to 31 °C, decreasing linearly to 50% relative humidity
at 40 °C
Mains supply ratings 100 – 240 V~, 50 – 60 Hz, 1.5 A
Mains supply voltage fluctuations not to exceed
±10% of the nominal voltage
Installation Category II (overvoltage categories)
Pollution Degree 2
See the maintenance section (Moving and Storage)
before moving the equipment.
Please note potential pinch hazards per the diagram
below (Figure C-1).
Additional hazards may exist if the equipment is not
used correctly per the user guide.