Section F – Operation
CD14 Tester & CD AutoPlus Autosampler User Guide—75-108-800 Rev. B—31 May 2021
78 of 130
EAR99 technology subject to restrictions on cover page.
NOTE: If “--.—” is displayed in any of the fields, that indicates
there is an issue with that temperature probe. Contact
Technical Support for assistance.
External Probe: The external temperature probe installed on
the dissolution tester.
Heater Control: The temperature probe located in the heater
which controls the rate at which the heater is heating the bath
water.
Calibrate Magazines (Cal Mag) tab
Touch the Cal Mag tab to access the magazine calibration
screen.
NOTE: This tab only appears if magazines are selected in the
configuration section.
Magazine Sample Travel: This field is the depth at which the
Magazine will travel to when the sample button is touched.
Magazine ADD Travel: This field is the distance in mm from
the home position to the ADD covers to implement a pill drop.
Home A/B: Homes the magazine by raising it up against the
drive head then moving it down a set amount.
Up: Moves the magazine up.
Sample: Moves the magazine to the sample position specified
in the Magazine Sample Travel field.
Pill Drop: Moves the magazine down to the height to trigger
Auto Dosage Delivery (ADD) covers.