Measurement Algorithms
High Speed Device Signal Quality Setup
To set up the equipment for the High Speed Signal Quality test, follow these
steps:
1.
Set the S6 switch to the Init position.
2.
Connect the standard USB cable between the Device SQ Init port and the
host port.
3.
Connect the A receptacle (marked DUT) from the Device SQ test port of the
test fixture to the unit under test (device).
4.
Configure the measurement and select the
command button to run the
application.
5.
Place the device in the test mode Test_Packet from the host controller.
6.
Set the S6 switch to the test position to isolate the unit under test while
maintaining the bus power.
7.
Select OK after acquiring a waveform. Verify that it is a correct waveform.
Figure 125: Equipment setup for High Speed Device testing
TDSUSB2 Universal Serial Bus Measurement Package
211
Содержание TDSUSB2
Страница 24: ...Getting Started 10 TDSUSB2 Universal Serial Bus Measurement Package...
Страница 32: ...Installation 18 TDSUSB2 Universal Serial Bus Measurement Package...
Страница 61: ...Operating Basics TDSUSB2 Universal Serial Bus Measurement Package 47...
Страница 76: ...Taking Measurements 62 TDSUSB2 Universal Serial Bus Measurement Package...
Страница 90: ...Tutorial 76 TDSUSB2 Universal Serial Bus Measurement Package...
Страница 160: ...Application Examples 146 TDSUSB2 Universal Serial Bus Measurement Package...
Страница 227: ...Measurement Algorithms Figure 126 Equipment setup for Chirp Test TDSUSB2 Universal Serial Bus Measurement Package 213...
Страница 230: ...Measurement Algorithms 216 TDSUSB2 Universal Serial Bus Measurement Package...
Страница 241: ...Measurement Algorithms Figure 138 Viewing SOF TDSUSB2 Universal Serial Bus Measurement Package 227...
Страница 250: ...Glossary 236 TDSUSB2 Universal Serial Bus Measurement Package...