Reference
Table 36.Error Code descriptions (cont.)
Error in acquiring waveform from oscilloscope
E111
Timeout occurred while
acquiring a waveform
This is the GPIB timeout information
that appears when the application is
acquiring the waveform
Make sure the probe and compliance test
fixture connections are proper. Reacquire
the new waveform
E112
Error in turning on a channel
The application failed to turn on the
live channel selected as source(s)
Make sure the probe and compliance test
fixture connections are proper. Reacquire
the new waveform
E113
Error in importing waveform
from instrument
The operation of importing the
waveform from the acquisition memory
failed. This will fail whenever there is
no valid waveform in the acquisition
memory
Make sure the probe and compliance test
fixture connections are proper. Reacquire
the new waveform
Packet Detection Warning
Try acquiring the signals again by pressing
Run button
E201
No EOP region found
Testing will be performed on an full
USB2.0 packet, which includes an
EOP region. If no EOP region is
found, the results might be incorrect
since testing will be performed on a
incomplete packet. For Upstream
signal quality testing, EOP must be
present
E202
Number of record points (data
points) are too few to process
At least 500 record points are required
to perform an operation. The
application automatically sets the
required record length. The user
should not change the record length
Try acquiring the signals again the
pressing Run button
E203
Idle region is not found after
EOP
As part of the USB2.0 packet
description, an EOP region follows an
idle region. If no idle region is found, it
may lead to incorrect results
Try acquiring the signals again by pressing
the Run button
E204
Reconfirm acquired data before
computing the results
As part of the USB2.0 packet
description, a minimum of three J-K or
K-J state transitions are expected. An
absence of these transitions may lead
to incorrect results
Try acquiring the signals again by pressing
the Run button
Signal rate measurement errors or warning
E301
There should be at least two
Cross-Over time values to
calculate the signal rate
For signal rate calculations, the
application expects a minimum of two
Cross-Over points
Try acquiring the signals again by pressing
the Run button
E306
Device/ File doesn't match with
the selected speed
The signal rate should be within 30%
of the specified signal rate of the Low,
Full and High Speed signals
Please check the speed of the connected
device. Ensure that the selection in the
application is the same. For example, if a
Low Speed unit is under test, make sure
Low Speed is selected in the application
170
TDSUSB2 Universal Serial Bus Measurement Package
Содержание TDSUSB2
Страница 24: ...Getting Started 10 TDSUSB2 Universal Serial Bus Measurement Package...
Страница 32: ...Installation 18 TDSUSB2 Universal Serial Bus Measurement Package...
Страница 61: ...Operating Basics TDSUSB2 Universal Serial Bus Measurement Package 47...
Страница 76: ...Taking Measurements 62 TDSUSB2 Universal Serial Bus Measurement Package...
Страница 90: ...Tutorial 76 TDSUSB2 Universal Serial Bus Measurement Package...
Страница 160: ...Application Examples 146 TDSUSB2 Universal Serial Bus Measurement Package...
Страница 227: ...Measurement Algorithms Figure 126 Equipment setup for Chirp Test TDSUSB2 Universal Serial Bus Measurement Package 213...
Страница 230: ...Measurement Algorithms 216 TDSUSB2 Universal Serial Bus Measurement Package...
Страница 241: ...Measurement Algorithms Figure 138 Viewing SOF TDSUSB2 Universal Serial Bus Measurement Package 227...
Страница 250: ...Glossary 236 TDSUSB2 Universal Serial Bus Measurement Package...