Running diagnostic tests
Running diagnostic tests
Use this procedure to test instrument functional modules.
NOTE.
Running self test resets the oscilloscope settings. Save the current setup to memory or a
fi
le if you do not want to
lose the current settings.
NOTE.
Remove all probes and cables from the oscilloscope inputs.
1.
Push the
Utility
button.
2.
Push the
-More- Page 1/2
side-menu button.
3.
Push the
Diagnostics
side-menu button.
4.
Push the
Self Test
side-menu button to display the
Self Test
menu and
Self Test Results
readout.
5.
Turn the
Multipurpose
knob to select
Run Self Test
, then click the
Multipurpose
knob to start testing using the
default settings. the self test takes a few seconds to run.
6.
Use the other
Self Test
menu items to re
fi
ne how to run the self test; you can run the tests a speci
fi
ed number of times
(
Loop Times
), run tests forever (
Loop Forever
), run tests until a test fails and stop testing (
Loop Until Fail)
, or run
a test on a failed tests (
Loop On Failure).
NOTE.
The oscilloscope resets to the factory default settings after running the self tests.
92
TBS2000 Series User Manual
Содержание TBS2072
Страница 1: ...x TBS2000 Series Oscilloscopes ZZZ User Manual P077114701 077 1147 01 ...
Страница 2: ......
Страница 18: ...Preface xii TBS2000 Series User Manual ...