Setting channel input parameters
Setting the probe attenuation factor
Attenuation is the amount that the probe reduces or ampli
fi
es the input signal amplitude before sending it to the oscilloscope
input. The
Attenuation
submenu sets the probe attenuation factor for probes that do not have the TekProbe II or TekVPI
interface. The default attenuation is
10x
.
1.
Push the
Vertical Menu
button for the
channel input that you want to set.
2.
Push the
Probe Setup
side menu to display the
Probe Setup
submenu.
3.
Use the
Multipurpose
knob to select and click the
Attenuation
submenu item.
4.
Use the
Multipurpose
knob to select and change the attenuation factor.
Quickly setting the probe attenuation to 1X or 10X
The
Set To 1X
and
Set To 10X
submenu items let you quickly set the probe attenuation to either 1X or 10X.
1.
Push the
Vertical Menu
button for the
channel input that you want to set.
2.
Push the
Probe Setup
side menu to display the
Probe Setup
submenu.
3.
Use the
Multipurpose
knob to highlight either
Set To 1X
or
Set To 10X
submenu item. Then click the multipurpose
knob to set that value.
Setting the Measure Current mode for voltage probes
If you are measuring current by using a standard voltage probe to measure the voltage drop across a resistor, use the
Measure Current
fi
eld to set the Amps/Volts or Volts/Amps ratio of your measurement setup. For example, if you are
measuring the voltage drop across a 2
Ω
resistor to determine the current, set the V/A ratio to 2.
1.
Push the
Vertical Menu
button for the
channel input that you want to set.
2.
Push the
Probe Setup
side menu to display the
Probe Setup
submenu.
32
TBS2000 Series User Manual
Содержание TBS2072
Страница 1: ...x TBS2000 Series Oscilloscopes ZZZ User Manual P077114701 077 1147 01 ...
Страница 2: ......
Страница 18: ...Preface xii TBS2000 Series User Manual ...