Analyzing a waveform
3.
Push the
Multipurpose
knob to select
the other cursor (which becomes a solid
line), then turn the knob to move that
cursor. The
fi
rst cursor is now drawn with
a dotted line.
4.
To make smaller cursor position
adjustments, push the
Fine
button to
toggle between making coarse and
fi
ne
cursor position adjustments.
The
Fine
button also enables smaller
adjustments on the
Vertical
and
Horizontal Position
knobs, the
Trigger
Level
knob, and many adjustment
operations of the
Multipurpose
knob.
5.
Push the
Amplitude
or
Screen
side-menu button to select a different cursor for taking measurements. (See page 57,
6.
Push the
Link
side-menu button to turn cursor linking
On
or
Off
. If linking is
On
, turning the
Multipurpose
knob
moves both cursors at the same time.
7.
If one or both of the cursors are off of the screen, push the
Bring Cursor On Screen
side-menu button to bring
off-screen cursors back to the screen.
8.
Push the
Cursors
front panel button to turn the cursors off.
Cursor types
The cursor types are:
Time or Frequency cursors. These vertical
cursors measure time or frequency, the
signal amplitude where the cursors cross the
waveform, and the absolute difference (delta)
between the two cursors cross points (both
time and amplitude delta).
Time readouts are relative to the trigger point
(which is 0 s). For example, cursors to the left
of the trigger are minus time values.
Amplitude cursors. These horizontal cursors
measure vertical amplitude parameters,
typically voltage.
TBS2000 Series User Manual
57