Page 2 of 2
TYPE 323
The charger and LOW BATT lamp circuit can be checked by first checking the
resistance of R615 and then performing Steps land 9 of the Calibration Procedure.
If circuit-drain and the charger have been exonerated as the source of
trouble by the preceding checks, the battery can be assumed to be at fault.
If
the probelm is a battery-operating time of shorter duration than that specified,
it may be due to a faulty cell(s) or to a general battery degradation caused by
abuse, old age or a high number of charge-discharge cycles.
In either case,
battery-operating time may be sufficiently long to satisfy operating demands,
and further troubleshooting can be ignored. However, if internal battery-
operating time is extremely short and/or rated internal battery-operating time
is to be regained, continue with this procedure.
To check the battery, proceed as follows:
(1) Complete a 24 hour FULL CHG cycle with the Oscilloscope PCWER switch OFF.
(2) Set the controls as specified for checking circuit-drain except that the
Power Pack switch should be set to FULL CHG and the external power should be
disconnected.
(3) Check the voltage between terminals M and I of the Battery Charger
circuit board to ascertain that no dead or shorted cells exist.
If the battery
voltage is below 7.2 V, individual cell voltages should be checked. To check
individual cells, remove the battery as explained in the Disassembly and Assembly
instructions in this section. Then use jumper wires to reconnect the battery
during this check, observing proper polarity. All cell voltages should be in
excess of 1.2 V.
Replace bad cells or the battery in accordance with the
information given in the Disassembly and Assembly instructions.
If individual
cells are replaced, recharge the battery for 24 hours before continuing with
this.procedure.
(4) Continue operating the Oscilloscope for approximately 2 hours. Again
check individual cell voltages. Any cell whose voltage is more than 50 mV below
the average voltage of the rest should be considered as suspect. Final deter
mination of its value can be made in step 5.
If individual cells are replaced,
recharge the battery for 24 hours before continuing with this procedure.
(5) Continue operating the Oscilloscope until the LOW BATT lamp blinks.
If
less than 4 hours time is obtained, the individual cell voltages should again be
checked against the information given in the preceding paragraph.
If individual
cell and average voltage are both low, the entire battery should be replaced.
If operating time is above 4 hours, cells or battery can be replaced at the
discretion of the customer.
Cll/169
Содержание 323
Страница 4: ...Type 323 Fig 1 1 Type 323 Oscilloscope ...
Страница 14: ...Operating Instructions Type 323 2 2 Fig 2 1 External controls connectors and indicators ...
Страница 39: ...Circuit Description Type 323 3 4 Fig 3 3 Paraphase Am plifier simplified ...
Страница 51: ...Circuit Description Type 323 3 16 Fig 3 8 Blocking Oscillator simplified ...
Страница 71: ...Maintenance Type 323 4 15 Fig 4 13 Transistor data ...
Страница 72: ...Maintenance Type 323 4 16 Fig 4 14 M ain circuit board p a rtia l loft side vertical circuit components ...
Страница 147: ...T Y P E 3 2 3 O S C I L L O S C O P E B L O C K D IA G R A M MRI4 i ...
Страница 157: ...BL OCK DIAGRAM ...
Страница 158: ......
Страница 161: ...A TYPE 323 OSCILLOSCOPE ...
Страница 162: ...1 TYPE 323 OSCILLOSCOPE ...
Страница 163: ...FIG 2 CABINET ...
Страница 164: ...OPTIONAL ACCESSORIES 016 0119 00 1 POWER PACK 016 0112 00 1 COVER protective oscilloscope ...