W AVEFORM com parisons should be m ade under the fo llo w in g
conditions except as noted on in d iv id u a l schematics:
Diagram Number
Equipment Setup
1, 2
Type 323 O scilloscope
V o lta g e
Source
A n y
V O LTS/D IV
5 DIV CAL
VARIABLE
CAL
T IM E /D IV
.1 ms
VARIABLE
CAL
POWER
O N
INTENSITY
M inim um intensity
consistent w ith
good vie w in g
POSITION
C ontrols
Knobs in and
d is p la y centered
TRIGGER
+ AUTO
T rig /H o riz
C oupling
INT T R I G - A C
POWER
O N
Test Oscilloscope
E xternally trig g e re d by positive
slope fro m Type 323 O scilloscope
CAL OUT
Sweep Rate
0 .5 m s/d iv is io n
3, 4, 6
Same as fo r dia gra m s 1 and 2,
except:
Type 323 O scilloscope T IM E /D IV at
50
j i
s
(C ontinued on diagram
\ 3 / )
Содержание 323
Страница 4: ...Type 323 Fig 1 1 Type 323 Oscilloscope ...
Страница 14: ...Operating Instructions Type 323 2 2 Fig 2 1 External controls connectors and indicators ...
Страница 39: ...Circuit Description Type 323 3 4 Fig 3 3 Paraphase Am plifier simplified ...
Страница 51: ...Circuit Description Type 323 3 16 Fig 3 8 Blocking Oscillator simplified ...
Страница 71: ...Maintenance Type 323 4 15 Fig 4 13 Transistor data ...
Страница 72: ...Maintenance Type 323 4 16 Fig 4 14 M ain circuit board p a rtia l loft side vertical circuit components ...
Страница 147: ...T Y P E 3 2 3 O S C I L L O S C O P E B L O C K D IA G R A M MRI4 i ...
Страница 157: ...BL OCK DIAGRAM ...
Страница 158: ......
Страница 161: ...A TYPE 323 OSCILLOSCOPE ...
Страница 162: ...1 TYPE 323 OSCILLOSCOPE ...
Страница 163: ...FIG 2 CABINET ...
Страница 164: ...OPTIONAL ACCESSORIES 016 0119 00 1 POWER PACK 016 0112 00 1 COVER protective oscilloscope ...