Calibration— Type 323
Fig. 6 -9 . Typical CRT d ispla y when checking in p u t capacitance.
c. Connect the probe tip to the BNC post jack.
d. Set the square-wave generator for a five-division display
at one kilohertz.
e. Compensate the probe as described in the probe instruc
tion manual.
f. CHECK—CRT display at each VOLTS/DIV switch set
ting for optimum square corner and flat top within + 3 % or
—3%, or total peak-to-peak aberrations not to exceed 3%.
Readjust the generator at each setting, remove the attenuator
and pull the X 10 VERT GAIN switch as given in Table 6-3
to maintain a five-division display.
g. ADJUST—VOLTS/DIV switch compensation as given in
Table 6-3. First adjust for optimum square corner on the dis
play and then for optimum flat top. Readjust the generator
output with each setting of the VOLTS/DIV switch, remove
the attenuator and pull the X 10 VERT GAIN switch as given
in Table 6-3 to provide five divisions of deflection. Fig.
6-1 OB shows the location of the variable capacitors.
h. Disconnect all test equipment and remove the calibra
tion shield.
25. Adjust Internal Trigger Compensation
O
a. Change the following control settings:
' M
7
E
V
* *
<0
Incorrect
top
ot fl a t)
MM
’
L
Fig. 6 -1 0 . (A ) Typical CRT d isp la y showing correct and incorrect
com pensation, (B) location o f com pensation adjustm ents (A tte n u a
tor b o a rd ).
VOLTS/DIV
.5
X 10 VERT GAIN
Pushed in
b. Connect the square-wave generator high-amplitude
output connector to the VERT INPUT connector through the
GR to BNC adapter, 42-inch 50-ohm BNC cable and 50-
ohm BNC termination.
c. Connect the 10X probe to the input connector of the
test oscilloscope.
d. Set the test oscilloscope for a vertical deflection fac
tor of 20 millivolts/division (0.2 volts/division at 10X probe
tip) at a sweep rate of 0.5 milliseconds/division. Adjust the
triggering controls when necessary to provide a stable dis
play.
®
6-15
Содержание 323
Страница 4: ...Type 323 Fig 1 1 Type 323 Oscilloscope ...
Страница 14: ...Operating Instructions Type 323 2 2 Fig 2 1 External controls connectors and indicators ...
Страница 39: ...Circuit Description Type 323 3 4 Fig 3 3 Paraphase Am plifier simplified ...
Страница 51: ...Circuit Description Type 323 3 16 Fig 3 8 Blocking Oscillator simplified ...
Страница 71: ...Maintenance Type 323 4 15 Fig 4 13 Transistor data ...
Страница 72: ...Maintenance Type 323 4 16 Fig 4 14 M ain circuit board p a rtia l loft side vertical circuit components ...
Страница 147: ...T Y P E 3 2 3 O S C I L L O S C O P E B L O C K D IA G R A M MRI4 i ...
Страница 157: ...BL OCK DIAGRAM ...
Страница 158: ......
Страница 161: ...A TYPE 323 OSCILLOSCOPE ...
Страница 162: ...1 TYPE 323 OSCILLOSCOPE ...
Страница 163: ...FIG 2 CABINET ...
Страница 164: ...OPTIONAL ACCESSORIES 016 0119 00 1 POWER PACK 016 0112 00 1 COVER protective oscilloscope ...