Calibration— Type 3 2 3
i. Gel lire TRIGGER control to — AUTO.
j. CHECK— Stable CRT display is presented with the Trig/
Horiz Coupling switch set to INT TRIG AC and ACLF REJ.
k. Set the constant-amplitude generator for a 0.75-divi
sion display at four megahertz.
l
. Pull the X 10 HORIZ MAG switch out.
m. CHECK—Stable CRT display is presented with the Trig/
Horiz Coupling switch set to INT TRIG AC and ACLF REJ.
n. Turn the TRIGGER control counterclockwise to the
variable negative-slope area.
o. CHECK— Stable CRT display can be obtained with the
Trig/Horiz Coupling switch set to INT TRIG AC and ACLF
REJ. TRIGGER control may be adjusted as necessary to ob
tain a stable display.
p. Turn the TRIGGER control counterclockwise to the vari
able positive-slope area.
q. CHECK—Stable CRT display can be obtained with the
Trig/Horiz Coupling switch set to INT TRIG AC and ACLF
REJ. TRIGGER control may be adjusted as necessary to
obtain a stable display.
r. Set the TRIGGER control to + AUTO.
s. CHECK— Stable CRT display is presented with the Trig/
Horiz Coupling switch set to INT TRIG AC and ACLF REJ.
47. Check External Triggering Operation
a. Change the following control settings:
Trig/Horiz Coupling
EXT TRIG OR
HORIZ AC
X 10 HORIZ MAG
Pushed in
b. Set the constant-amplitude generator for a 0.75-divi
sion display (75 millivolts) at 400 kilohertz.
c. CHECK—Stable CRT display is presented.
d. Turn the TRIGGER control clockwise to the variable
positive-slope area.
e. CHECK— Stable CRT display can be obtained with the
Trig/Horiz Coupling switch set to EXT TRIG OR HORIZ AC
and DC. TRIGGER control may be adjusted as necessary
to obtain a stable display.
f. Turn the TRIGGER control clockwise to the variable
negative-slope area.
g. CHECK— Stable CRT display can be obtained with the
Trig/Horiz Coupling switch set to EXT TRIG OR HORIZ AC
and DC. TRIGGER control may be adjusted as necessary
to obtain a stable display.
h. Set the TRIGGER control to — AUTO.
i. CHECK— Stable CRT display is presented.
j. Set the constant-amplitude generator for a 1.9-division
display (190 millivolts) at 400 kilohertz.
k. Without changing Ihe oUlpUl amplitude, set the yell
erator to four megahertz.
l
. Pull the X 10 HORIZ MAG switch out.
m. CHECK— Stable CRT display is presented.
n. Turn the TRIGGER control counterclockwise to the vari
able negative-slope area.
o. CHECK— Stable CRT display can be obtained with the
Trig/Horiz Coupling switch set to EXT TRIG OR HORIZ AC
and DC. TRIGGER control may be adjusted as necessary
to obtain a stable display.
p. Turn the TRIGGER control counterclockwise to the vari
able positive-slope area.
q. CHECK— Stable CRT display can be obtained with the
Trig/Horiz Coupling switch set to EXT TRIG OR HORIZ AC
and DC. TRIGGER control may be adjusted as necessary to
obtain a stable display.
r. Set the TRIGGER control to + AUTO.
s. CHECK— Stable CRT display is presented.
t. Disconnect the high-frequency generator.
48. Check Low-Frequency Triggering Operation
a. Connect the low-frequency constant-amplitude sine-
wave generator to the VERT INPUT connector through the
42-inch 50-ohm BNC cable, 50-ohm BNC termination and the
BNC T connector. Connect the output of the BNC T con
nector to the EXT TRIG OR HORIZ INPUT connector with a
42-inch 50-ohm BNC cable.
b. Change the following control settings:
TIME/DIV
10 ms
X 10 HORIZ MAG
Pushed in
c. Set the low-frequency generator for a 0.75-division
display (75 millivolts) at 30 hertz.
d. CHECK— Stable CRT display is presented.
e. Turn the TRIGGER control clockwise to the variable
positive-slope area.
f. CHECK— Stable CRT display can be obtained with the
Trig/Horiz Coupling switch set to EXT TRIG OR HORIZ
AC and DC. TRIGGER control may be adjusted as neces
sary to obtain a stable display.
g. Turn the TRIGGER control clockwise to the variable
negative-slope area.
h. CHECK— Stable CRT display can be obtained with the
Trig/Horiz Coupling switch set to EXT TRIG OR HORI2!, AC
and DC. TRIGGER Control may be adjusted as necessary to
obtain a stable display.
i. Set the TRIGGER control to — AUTO.
j. CHECK— Stable CRT display is presented.
k. Set the Trig/Horiz Coupling switch to INT TRIG AC.
6-26
® I
Содержание 323
Страница 4: ...Type 323 Fig 1 1 Type 323 Oscilloscope ...
Страница 14: ...Operating Instructions Type 323 2 2 Fig 2 1 External controls connectors and indicators ...
Страница 39: ...Circuit Description Type 323 3 4 Fig 3 3 Paraphase Am plifier simplified ...
Страница 51: ...Circuit Description Type 323 3 16 Fig 3 8 Blocking Oscillator simplified ...
Страница 71: ...Maintenance Type 323 4 15 Fig 4 13 Transistor data ...
Страница 72: ...Maintenance Type 323 4 16 Fig 4 14 M ain circuit board p a rtia l loft side vertical circuit components ...
Страница 147: ...T Y P E 3 2 3 O S C I L L O S C O P E B L O C K D IA G R A M MRI4 i ...
Страница 157: ...BL OCK DIAGRAM ...
Страница 158: ......
Страница 161: ...A TYPE 323 OSCILLOSCOPE ...
Страница 162: ...1 TYPE 323 OSCILLOSCOPE ...
Страница 163: ...FIG 2 CABINET ...
Страница 164: ...OPTIONAL ACCESSORIES 016 0119 00 1 POWER PACK 016 0112 00 1 COVER protective oscilloscope ...