24
Operation
To set the temperature controller to the required operating temperature, switch on the electrical
supply which will illuminate the power and cut-out indicators. Then set the required temperature,
by using the up and down arrows on the Eurotherm controller. Depress the reset switch to initiate
the system and extinguish the cut-out indicator.
The following parameters have been factory set into the 2208L controller:
Al List
Hy =
1
Pid List
Pb =
20
ti =
360
td =
60
lcb =
Auto
Hcb =
Auto
OP List
OpLo =
0
OpHi =
100.0
CycH =
1.0
OntH =
0.1
Accs List (Full level)
Enter 5744 password
Goto “OPer” select “Full”
Enter
SP List
SSEL =
SP1
SP1 =
0
SP2 =
25
SP1L =
0
SP1H =
630
SP2L =
0
SP2H =
630
SP1r =
Off
Dwell =
Off
endt =
rset
Prog =
rset
Stat =
dwell
IP List
filt =
1.6
OfSt =
-20
CjCo =
40
Mv =
19.7
For older units the following control parameters have been factory pre-set on the Eurotherm 807:-
High Alarm
HiAl
630ºC
1166ºF
Low Alarm
Lo Al
0ºC
32ºF
Deviation Alarm
D. Al
630ºC
1166ºF
Proportional Band
P
17%
17%
Integral Time
I
124 sec
124 sec
Derivative Time
D
3 sec
3 sec