background image

 

Safety and Installation 

 
Please read all the information in this booklet before using the unit. 

WARNING 

HIGH  TEMPERATURES  ARE  DANGEROUS:  they  can  cause  serious  burns  to  operators  and 
ignite combustible material. 

Techne have taken great care in the design of these units to protect operators from hazards, but 
Operators should pay attention to the following points: 

• USE CARE AND WEAR PROTECTIVE GLOVES TO PROTECT HANDS. 
• DO NOT put hot objects on or near combustible objects. 
• DO NOT operate the unit close to inflammable liquids or gases. 
• DO NOT place any liquid directly in your unit. 
• At all times USE COMMON SENSE. 

OPERATOR SAFETY 

All users of Techne equipment must have available the relevant literature needed to ensure their 
safety.  It is important that only suitably trained personnel operate this equipment, in accordance 
with the instructions contained in this manual and with general safety standards and procedures. 
If  the  equipment  is  used  in  a  manner  not  specified  by  Techne  the  protection  provided  by  the 
equipment to the Operator may be impaired. 

All  Techne  units  have  been  designed  to  conform  to  international  safety  requirements  and  are 
fitted  with  an  overtemperature  cutout. On  some models,  the  cutout  is  adjustable  and  should  be 
set to suit the application. On all other models the cutout is preset to protect the unit. 

If a safety problem should be encountered, the unit should be switched off and isolated from the 
mains electricity supply. 

INSTALLATION 

This  Techne  Industrial  Fluidised  Bath  (IFB)  is  designed  to  be  operated  as  a  permanently 
connected  piece  of  equipment  and  must  be  installed  by  a  fully  qualified  electrician  with 
experience of installing 3 phase equipment. 

The IFB must be installed with an appropriately rated switch and/or circuit breaker included in the 
building installation. 

The switch or circuit breaker should be in close proximity to the IFB and within easy reach of the 
operator. 

The  switch  or  circuit  breaker  should  also  be  clearly marked  as  the  disconnecting  device for the 
IFB. 

The supply wiring should be of a specification suitable for the ratings indicated on the IFB rating 
plate;  in  addition  it  should  also  comply  with  local,  national  and  international  standards  where 
applicable. 

As every installation is unique, it is not possible to give detailed instruction regarding any drilling 
of the cabinet base plate that may be required for routing of the supply cable. However, the cable 
must  be  adequately  secured  with  a  suitably  approved  cable  gland  at  the  point  it  enters  the 
cabinet.  It  is  also  a  requirement  that  the  earth  conductor  is  terminated  in  such  a  way  that  it  is 
longer than the 3 phase conductors and also longer than the neutral conductor. This is to prevent, 
in the case of the supply cable becoming insecure, the earth conductor disconnecting before the 
other 4 conductors. 

Содержание IFB-201

Страница 1: ...IFB 201 OPERATOR S MANUAL...

Страница 2: ...ERS 10 INSTALLATION 10 NACH DEM GEBRAUCH 11 GARANTIE 11 WARTUNG DURCH DEN BEDIENER 11 INTRODUCCI N 13 AVISO 13 INSTALACI N 13 DESPU S DE SU USO 14 GARANT A 14 MANTENIMIENTO 14 INTRODUZIONE 16 ATTENZIO...

Страница 3: ...3 Spare Parts 31 Accessories 31 Circuit Diagrams 32...

Страница 4: ...protect the unit If a safety problem should be encountered the unit should be switched off and isolated from the mains electricity supply INSTALLATION This Techne Industrial Fluidised Bath IFB is desi...

Страница 5: ...Y PROPERLY TRAINED PERSONNEL REMOVING THE SIDE FRONT OR REAR PANELS EXPOSES POTENTIALLY LETHAL MAINS VOLTAGES THERE ARE NO OPERATOR MAINTAINABLE PARTS WITHIN THE EQUIPMENT In the unlikely event that y...

Страница 6: ...serait alors plus garantie Tous les appareils Techne sont con us pour r pondre aux normes de s curit internationales et sont dotes d un coupe circuit en cas d exc s de temp rature Sur certains mod les...

Страница 7: ...rantie condition ou repr sentation concernant la description commercialisation adaptation aux conditions d utilisation ou autre des appareils ou leurs composants ENTRETIEN UTILISATEUR IMPORTANT CET AP...

Страница 8: ...votre fournisseur pour r paration 3 Fusibles La protection de l appareil est assur e par un ou deux fusibles dont le remplacement ne peut tre effectu que par un personnel qualifi Si les fusibles saute...

Страница 9: ...die pers nliche Sicherheit des Anwenders beeintr chtigen Die Ger te von Techne entsprechen den internationalen Sicherheitsbestimmungen und sind mit einem automatischen bertemperaturabschalter ausgest...

Страница 10: ...lich dazu erstellt um Kunden die Handhabung der Techne Ger te zu erleichtern Nichts in dieser Gebrauchsanleitung darf als Garantie Bedingung oder Voraussetzung verstanden werden sei es die Beschreibun...

Страница 11: ...zur ck indem Sie den roten Knopf einmal bis zum Anschlag dr cken Wenn der bertemperaturabschalter wiederholt ausl st liegt ein gr erer Defekt vor Das Ger t mu zur Reparatur an Ihren Lieferanten einges...

Страница 12: ...rporados un sistema de desconexi n en caso de sobretemperatura En algunos modelos el sistema de desconexi n es variable lo que le permite elegir la temperatura seg n sus necesidades En otros el sistem...

Страница 13: ...garant a condici n o aseveraci n de la descripci n comerciabilidad o adecuaci n para cualquier fin espec fico de los aparatos o sus componentes MANTENIMIENTO ESTE APARATO DEBE SER DESMONTADO SOLO Y EX...

Страница 14: ...a desconexi n repetida indicar a una aver a grave puede que tenga que devolverle el aparato a su proveedor para su reparaci n 3 Fusibles Su aparato est protegido por uno o dos fusibles S lo deben camb...

Страница 15: ...di sicurezza e sono equipaggiate con un interruttore anti surriscaldamento Su alcuni modelli l interruttore regolabile e dovrebbe essere impostato secondo l utilizzo In tutti gli altri modelli l inte...

Страница 16: ...zione dei pannelli laterali frontali o posteriori pu esporre potenzialmente a voltaggi di corrente letali All interno dell apparecchio non ci sono parti manutenibili da parte dell utilizzatore Nell ev...

Страница 17: ...n of inflammable fumes above the bath and within the exhaust duct work must be kept below the lower explosive limit General Description The fluidised bath employs the principle of fluidisation of a ma...

Страница 18: ...sing air control valves Supplied as standard with this and all other options is an air pressure regulator and filter assembly During operation the correct fluidising air flow rate must be determined b...

Страница 19: ...80 minutes 20 C to 600 C 350 minutes Compressed air supply pressure 204 1030kPa 30 150psi Air consumption maximum 220 l min At 450 C 150 l min Nominal Heater Power 9kW Heater type 4 semi embedded heat...

Страница 20: ...nce between the top of the aluminium oxide and the top of the bath is 120mm when the bath is fully loaded Should the fluidised bath be stored for long periods of time under damp or humid conditions mo...

Страница 21: ...that the maximum fume concentration be less than 25 of the LEL f Measure the fume concentration in the working environment adjacent to the fluidised bath using suitable gas analysis equipment such as...

Страница 22: ...rticle size less than 0 005mm diameter which will not be retained by the cyclone it is recommended that a filtration system be fitted after the cyclone In applications where the materials being treate...

Страница 23: ...8L controller Al List Hy 1 Pid List Pb 20 ti 360 td 60 lcb Auto Hcb Auto OP List OpLo 0 OpHi 100 0 CycH 1 0 OntH 0 1 Accs List Full level Enter 5744 password Goto OPer select Full Enter SP List SSEL S...

Страница 24: ...luidised bath We recommend that protective clothing gloves aprons and visors etc be worn at all times when working with fluidised baths and that the installation and maintenance procedures outlined in...

Страница 25: ...d they should be allowed to cool in air and whilst still warm treated to prevent rusting It should be noted that some plastics and in particular paints contain fillers These fillers are usually inorga...

Страница 26: ...bath is self draining However it should be kept in good condition by inspection at two weekly intervals and by cleaning the bowls and washing or replacing the filter elements as necessary With excepti...

Страница 27: ...et In addition in water it produces water ions which even in neutral or alkaline solutions promote corrosion and rusting in steel witness for example the corrosive nature of sea water and calcium chlo...

Страница 28: ...le the medium may be thoroughly washed in several changes of clean water washed through a 70 mesh sieve drained dried and returned to the bath where drying is completed by fluidisation and heating at...

Страница 29: ...ally terminated within the control console see attached circuit diagrams for wiring details To inspect the heater elements remove the inner container from the outer wrapper The heater elements are of...

Страница 30: ...457 6500356 6500356 6500356 Fuse 2 Amp 6008321 6008321 6008321 6008321 6008321 Fuse 16 Amp N A N A 6008322 6008322 6008322 Fuse 35 Amp 6008324 6008324 N A N A N A Filter Kit Oil 6009247 6009247 600924...

Страница 31: ...32 Circuit Diagrams...

Страница 32: ...33...

Страница 33: ...34...

Страница 34: ...g the details below UK Office Keison Products P O Box 2124 Chelmsford Essex CM1 3UP England Tel 44 0 330 088 0560 Fax 44 0 1245 808399 Email sales keison co uk Please note Product designs and specific...