59
2.5nm, 1nm and closed.
PMT Mode and PMT Value
PMT Mode refers to the boosting mode of the photomultiplier. PMT Value can be entered in
manual mode, the range is 0
~
900V.
Open the cover plate of the sample room, the high voltage will automatically switch off.
Close the cover plate and it is necessary to reset in the interface of “Setting”.
3)
Quantitation
Figure 3-52 Dialog box of parameters for Quantitation
EX WL and EM WL can be selected from 200nm to 900nm;
Calibration: Standard sample fitting, “Manual Fitting” can be selected, fitting method:
1
st
order, 2
nd
order, 3
rd
order and Segmented.
Fitting Coefficient: A0, A1, A2 and A3;
“Force Zero” can be selected (machine fitting).
Replicates: Standard 1~100;
Sample 1~100.
Sample Concentration: Lower limit and upper limit can be set and concentration unit
can also be entered manually.