1.4.3 Precautions on Using Workstation
This workstation is an extension from the computer and apply advanced software to
manage the whole system. The work status of the instrument is complex and the data volume is
large, which result in heavy loading on the computer. Thus, the computer with the workstation
should never be used as a general computer. Games playing on this computer is not allowed.
Otherwise, if the workstation is infected by computer viruses, the consequences will be
unimaginable, and Techcomp and its representative is not responsible for the damage done by
this cause, even within the warranty period.
1.5 Xenon Lamp Installation and Adjustment
During the installation of xenon lamp, in order to prevent electric shock, make sure the AC
power cord is not connected to the power supply.
Please follow below procedure for installation:
1
)
Remove 2 screws a on the outer cover of the lamp compartment
①
and
detach the outer cover from the instrument (Figure 1-2). Then disconnect the
socket to the fan
②
(Figure 1-3);
2
)
Remove 3 screws
④
on the inner cover of the lamp compartment and detach
the inner cover
③
from the lamp compartment (Figure 1-4);
3
)
Profile of xenon lamp for reference (Figure 1-5);
4
)
The light source safety switch on the inner cover of the light source is shown
in Figure 1-6. It can be changed to Figure 1-7 by unscrewing 2 screws;
5
)
Install the inner cover of the lamp compartment. Be careful to press the light
source safety Switch (Figure 1-8);
6
)
Tighten 3 screws on the inner cover and return them to the status shown in
Figure 1-4. At this stage, turn on the power supply and check and optimize the
energy of the lamp according to Chapter 1.6. After optimizing the lamp
energy, connect the power cord of the fan
②
on the outer cover of the lamp
compartment (Figure 1-3). Install the outer cover of the lamp compartment
with 2 screws to the status shown in Figure 1-2. Turn on the instrument and
start using it.