1.3 Instrument Connection
Figure 1-1 Instrument constitution
1. Display 2. Computer 3. Keyboard 4. Host 5. Printer 6. Mouse
1.4 Installation Site: Installation and Adjustment
1.4.1 Installation Site
In order to full utilize the instrument and to keep performance for long term usage, please
install the instrument in the sites that meet the following requirements. Techcomp and its
representative will not be responsible for the performance deterioration and damage of the
instrument due to operate the instrument in the sites that do not meet the specified requirements,
even within the warranty period. Please pay attention to it.
1) The room temperature should be between 10 ℃ and 30 ℃ when we measuring the
samples;
2) Avoid direct sunlight;
3) Strong vibration or continuous weak vibration enough to be felt by the human body
must not be transmitted to the instrument;
4) Avoid strong magnetic field or strong electric field. There should be no equipment
that can generate high frequency wave nearby;
5) The humidity should not exceed 85%;
6) There should be no corrosive gas, organic and inorganic gases that there is no
absorption in the region of UV
7) Room should be with less dust;
8) The stability of the power supply should meet the requirements of the instrument.
1.4.2 Power
1) The power supply voltage of FL970 should be 220V/50Hz, fluctuation within
±
10%
is allowed, the fuse is 3A. It is recommended to configure the on-line stabilized
voltage power supply or UPS to keep the instrument work normally all the time.
2)
The instrument is controlled and the data is processed by the computer. Please do not
share the same power supply with the other equipment that can generate strong pulse
signal.