8. Maintenance and Cleaning
68
HYDROFLEX PLUS No. 30199906 Rev. No. 1.1
2022-01
8.6.2
Removing the Manifold
WARNING
PRIME WITHOUT LIQUID TO REMOVE ANY LIQUIDS FROM THE
TUBING SYSTEM.
The manifold should be removed and thoroughly cleaned at least once every
6 months or whenever needles become blocked.
1. Lift the manifold arm and remove the black manifold arm guard plate by sliding
it out from behind the three screws
– the screws do not need to be removed to
perform this step.
2. Lift the manifold arm and remove the two screws that attach the manifold to
the instrument using the Allen key provided.
3. Carefully pull the tubing off the connectors on the rear of the manifold and
remove the manifold.
3 screws
2 screws