6. Settings Menu
2022-01
HYDROFLEX PLUS No. 30199906 Rev. No. 1.1
49
6. Settings Menu
6.1
Introduction
The
Settings
menu has the following options:
Edit Plates
Adjustment procedure for setting the parameters for up to
20 different microplates.
Options
Depending on available options (e.g.,
Edit LLD Channels
).
Bubble Sensor
The
Bubble Sensor
sensitivity can be set to
Low,
Medium
, or
High
depending on liquids used (if this option
is installed). If very foamy liquids are used the bubble
sensor should be switched
OFF
.
The
Settings
menu has the following structure:
EDIT PLATES
< > EXIT OK
SETTINGS
< > OK
OPTIONS
< > EXIT OK
BUBBLE SENSOR
< > EXIT OK
6.1.1
Edit Plates
EDIT PLATES
- + EXIT OK
Plate 1, GRE96ft
- + EXIT OK
Z-User Def 14600
- + EXIT OK
OVERFLOW POS.
< > EXIT OK
Press OK to continue.
Press
– or + to select the plate (1-20)
and press OK to edit the plate parameters.
Use the cursor keys to select the position to be defined
(OVERFLOW, BOTTOM, ASP1 or ASP2) and press OK.
For round and V-shaped bottom plates,
only one aspiration can be defined.
The instrument moves to the predefined position.
Use the
– and + keys to change the position. Confirm with OK.
The next position can then be selected.
Press OK to save the defined position; Press EXIT to leave the
menu without saving the position. The instrument moves to the
home position and the firmware returns to the Edit Plates menu.
EDIT PLATES
- + EXIT OK