2. General
2022-01
HYDROFLEX PLUS No. 30199906 Rev. No. 1.1
19
PARAMETERS
CHARACTERISTICS
Dispensing uniformity
<= 4 % CV
Measured under following conditions:
8-way manifold, 300 µl, dispensing rate 3 (250 µl/s/well), wash buffer, Greiner
96 well flat bottom plate
Residual volume
Measured under following conditions:
<= 2 µl / well (Greiner 96-well flat bottom plate)
-
Crosswise aspiration, 8-way manifold, aspiration rate: 3, aspiration time: 4 s,
head speed: 10 mm/s, wash buffer.
-
Residual volume of <= 2 µl / well cannot be guaranteed if a 16-way manifold is
used and if not, all wells are filled with liquid.
<= 5 µl / well (Greiner 96-well round bottom and V-shaped bottom plates)
-
Single aspiration, 8-way manifold, aspiration rate: 3, aspiration time: 4 s, head
speed: 10 mm/s, wash buffer.
-
Residual volume of <= 5 µl / well cannot be guaranteed if a 16-way manifold is
used and if not, all wells are filled with liquid.
Carry over
between wells
<= 1 ppm
The Carry over between wells <= 1 ppm cannot be guaranteed if not all wells
are filled with liquid.
Power
Supply
Auto-sensing
100 - 120 V or 220 - 240 V, 50/60 Hz
Consumption
< 65 VA
Physical
Outside dimensions
Width: 275 mm, Depth: 366 mm, Height: 180.5 mm
(Width: 10.8 inches, Depth: 14.4 inches, Height: 7.1 inches)
Weight
6.6 kg
Environmental
Ambient temperature
Operation
15°C to 35°C (59°F to 95°F)
Storage
-20°C to 60°C (-4°F to 140°F)
Relative Humidity
Operation
20 % to 80 %
non-condensing
Storage
10 % to 80 % non-condensing
Others
Overvoltage category
II
Pollution degree
2
Method of disposal
Electronic waste (infectious waste)