2. General
24
HYDROFLEX PLUS No. 30199906 Rev. No. 1.1
2022-01
2.5.2
Fitting and Connecting the Liquid Level Detection (LLD)
(Optional)
Connect the liquid sensor cables into the appropriate sockets in the rear panel of
the instrument. The Liquid Level Detection system can be extended to control up
to four different liquid bottles and one waste bottle.
Fig. 2.1 Waste and Liquid Bottles with LLD Sensors
See chapter 3.3.1 Connection Diagram for further details.
2.6
Microplate Requirements
Only 96-well format microplates (round, v-shaped, and flat bottom - including strip
plates) can be used with the HYDROFLEX PLUS.
PARAMETERS
CHARACTERISTICS
Max. overall plate height
14.35 mm ± 0.76 mm
(0.5650 inches ± 0.0299 inches)
Footprint (ANSI/SBS 1-2004)
127.76 mm x 85.48 mm
(5.0299 in x 3.3654 in)
Pitch size (center to center)
9.0 mm (0.3543 inches)
Bottom shape
Round, v-shaped, and flat