8. Maintenance and Cleaning
2022-01
HYDROFLEX PLUS No. 30199906 Rev. No. 1.1
61
Priming is performed using the following procedure:
PRIME
< > OK
CHANNEL 1
- + EXIT OK
WASTE BOTTLE OK?
< > YES
PRIME SOLUTION OK?
< > YES
Confirm rinsing solution and tube in bottle
Select and confirm PRIME procedure
Select required channel
(1 - 4, depending on instrument configuration)
Select OK and the prime procedure starts
Confirm waste bottle is empty
PRIME
< > OK
PROGRAM
< > OK
Instrument is ready to perform wash procedures
TIME
< > OK
VOLUME
< > OK
Select the duration TIME (5
– 99 s) or the
VOLUME (15
– 820 ml)