Communication GUI Guide
GXE600 Series
TDK-Lambda
<Page>
38/41
#
Starting Address
Qty. Function
8
Over Current Threshold
106
1
Register for setting overcurrent threshold.
Specify SWOCP detection threshold with Q10 format data
normalized with maximum current.
T he register setting value
Max.
Min.
Factory setting
Unit
Q10 format data
1.2
0.2
1.2
-
(Integer notation)
1229
204
1229
-
9
PV Rising T ime
107
1
Register for setting the PV reference slew rate at the start of output.
Specify the PV reference slew rate (dv / dt) at the start of output.
T he register setting value
Max.
Min.
Factory setting
Unit
Integer data
10000
20
20
ms
10
CC Rising T ime
108
1
Register for setting the CC reference slew rate at the start of output.
Specify the CC reference slew rate (di / dt) at the start of output.
T he register setting value
Max.
Min.
Factory setting
Unit
Integer data
10000
20
20
ms
11
PV Transition Time
109
1
Register for setting PV reference value slew rate at steady output.
Specify the PV command slew rate (dv / dt) at steady output.
T he register setting value
Max.
Min.
Factory setting
Unit
Integer data
10000
20
20
ms
12
CC Transition Time
110
1
Register for setting CC reference value slew rate at steady output.
Specify the CC command slew rate (di / dt) at steady output.
T he register setting value
Max.
Min.
Factory setting
Unit
Integer data
10000
20
20
ms
13
PF Threshold
111
1
Register for setting PF (Power Fail) detection threshold.
Specify the detection threshold when the PF signal is output.
The reference value is selected by setting the PF mode register.
T he register setting value
Max.
Min.
Factory setting
Unit
Integer data
95
50
80
%
14
PF Mode
112
1
Register for setting PF detection mode.
Specify the detection target (output voltage or output current) of PF.
CV Mode (Value 0) : When the output voltage drops below
the PF threshold value, the PF signal is output.
CC Mode (Value 1) : When the output current drops below
the PF threshold value, the PF signal is output.
T he register setting value
CV PF
CC PF
Factory setting
Unit
Integer data
1
0
0 (CV)
-
Data form
(0x0070)
Data form
Data form
(0x006D)
Data form
(0x006E)
Data form
(0x006F)
(0x006C)
Holding Register
(0x006A)
Data form
(0x006B)
Data form