Note
- For more details about these values, please see the User's Manual that came
with your sensor.
[Get Settings] Button
This allows you to gather the equation that is currently set in the RVR-
52
A
logger.
[Send Settings] Button
This sends the currently displayed moisture percentage calculation
equation to the RVR-
52
A logger.
- Moisture Adjustment
Before making adjustments, use the [Get Settings] Button to receive the
current settings.
By placing a check next to [Make Adjustments], it is possible to make
corrections for measurement errors that may occur.
This function allows you correct or adjust for errors in measurement values.
The Data Logger will record only the adjusted value.
Current Reading: Enter the actual measurement before adjustment
Post-adjusted Reading : Enter the desired measurement after adjustment.
[Use Adjustment in Equation] Button
The newly adjusted equation will appear in the Moisture Percentage
Calculation Equation Area.
[Send Settings] Button
The adjusted equation will be sent to the Data Logger.
Note
- When you are carrying out adjustments and formulating equations please take
care with the settings and take personal responsibility for the process.
- We cannot guarantee that after carrying out adjustment the measuring accuracy
will improve for all measuring ranges.
- For details about Moisture Adjustment, see the [Help Menu] in the provided
software T&D Recorder for Windows.