Pt: Value Adjustment Function
This function enables the user to make adjustments to measured
values to correct inaccuracies that may occur when compared to a
standard or reference temperature measurement. By setting an
adjustment equation beforehand, the adjustment function will
automatically calculate to record and display the post-adjusted
measurement. It is possible to select from two methods of
adjustment:
1
-point adjustment and
2
-point adjustment.
Note
- By changing the sensor, the current recording session will be automatically stop
■
Temperature:
1
-point Adjustment
Adjustment is carried out by entering the value of the difference between the
displayed temperature measurement reading and the reference one to which
it is being compared.
- Please set so that the possible adjustment range is within
±
54
°
F.