52
3
.
Select either [
1
-point Adjustment] or [
2
Point Adjustment].
4
.
Please enter the actual pre-adjusted value(s) in the [Current Reading]
column and the post-adjusted reference value(s) in the [Post-adjusted
Reading] column.
5
.
Click on the [Send Settings] Button to send the adjustment equation to
the selected Data Logger.
[Initialize] Button
Clears any previous adjustment equation(s) from the Data Logger.
[Send Settings] Button
Transmits the adjustment equation to the Data Logger.
[Close] Button
Quits and closes the window.
Note
- We cannot guarantee that after carrying out adjustment the measuring accuracy will
improve for all measuring ranges.
- By carrying out adjustment on an already adjusted device without first clicking the
Initialize Button, the new adjustment equation will be be used to further adjust the
previously adjusted result. It is important to remember that if adjustment takes place
without initializing the device, the adjustment will include the first equation as well as
the second and an over adjustment will occur. This over adjustment will be
especially large when making a
2
-point adjustment and the measurement falls
outside of the range of the two points entered.
When you are carrying out adjustments and formulating equations please take care
with the settings and take personal responsibility for the process.