Temperature:
2
-point Adjustment
Adjustment is carried out by entering values for the differences between two
displayed temperature measurement readings and the reference ones to
which they are being compared.
- Please set so that the possible adjustment range is within
±
5
°
F.
■
Adjustment Settings
1
.
Please follow directions as found in [Communication between Data
Loggers and Computers] by connecting the RTR-
57
U unit to your
computer and place the Data Logger unit face down on top of it.
(see p.
21
for details)
2
.
In the Main Window, click on the Data Logger icon. To open the
[Adjustment (RTR-
51
/
52
)] or [Adjustment (RTR-
52
Pt)] Window, click on
[Settings] in the Menu Bar and then select [RTR-
51
/
52
Adjustment
Setting] or [RTR-
52
Pt Adjustment Setting] from the pull down menu.
[Initialize] Button
[Send Settings]
Button
Enter current and post-adjusted reading(s)
Select
1
-point
Adjustment or
2
-point
Adjustment