69
■
Moisture Probe Settings
If under the [Start Recording] Tab, you have selected "Measure
Moisture by Percentage" as the Measuring Mode, then here you
should set into the Data Logger the moisture percentage calculation
equation for the moisture sensor you are using.
1
.
Please follow directions as found in [Communication between Data
Loggers and Computers] by connecting the RTR-
57
C/
57
U unit to your
computer and place the RVR-
52
A unit face down on top of it. (see P.
21
for details)
2
.
In the Main Window, click on the RVR-
52
A icon. To open the [Moisture
Probe Settings] Window, click on [Settings] in the Menu Bar and then
select [Moisture Probe Settings] from the pull down menu.
Moisture
Adjustment
Probe
Settings
[Use Adjustment in Equation]
Button
[Send Settings]
Button
[Get Settings]
Button
- Probe Settings
Select which sensor you are using, either EC-
10
or EC-
20
and click the
[EC-xx Default Values] Button and [Send Settings] Button for that sensor.
- EC-
10
Default Values: Y =
0
.
0936
X -
37
.
6
- EC-
20
Default Values: Y =
0
.
0695
X -
29