A:
Check to see if some other communication software is in use.
A:
Most desktop models have two serial ports, try changing the plug to
the other port and change the settings.
Q:
The date and the time of the recorded data are different from
the actual date and time. What should I do?
A:
The Thermo Recorder takes the date and time from your computer
when it is set up. Make sure that the date and time are correct on
your computer.
Q:
The date and the time of the recorded data are different from
the actual date and time of recording. Why?
A:
When recording mode has been set to "Endless" and Recording
Start was carried out via PC, the time for recorded data is calculated
from the recording start date and time and will appear correctly in
the Graph as long as recording capacity has not reached its limit.
However, once data capacity has become full and the oldest data is
overwritten, the date and time for downloaded data in the Graph will
be calculated using the PC's clock for the latest recorded data and
subtracting backward. Hence the time for the displayed data will be
different from the actual time the data was recorded.