* Menu item "Low alarm"
Optional: min SP to max SP. resolution 0.1°C. No
hardware connected, display function only.
"Menu item "High alarm"
Optional: min SP to max SP. resolution 0.1°C. No
hardware connected, display function only.
* Optional
Menu item "PID parameter"
PID set 1
-
First set of parameters
PID set 2
-
Second set of parameters
PID set 3
-
Third set of parameters
PID set 4
-
Activated when communication
via RS232
Each set offers individual PID settings for:
Proportional band value
Integrating value
Differentiating value
Default settings
Pb*
:=
25
I
:=
16
D
:=
0
*Pb= proportional band. Proportional value P is found to be 100/Pb.
Menu item "Backlight"
On
Off
Standard value: On
Menu item "Temp units"
°C
°F
Standard value: °C
Menu item "Baudrate"
300
600
1200
2400
4800
9600
19200
38400
Standard value: 9600