4.6
Maximum Level
If the fluid level is too high, it will leave the bath via the
overflow outlet (10mm outside diameter pipe). Prevent
fluid from the overflow outlet entering the backside of the
TV7000. For this reason the overflow outlet must be
connected to a waste container.
When the bath is working at high temperatures, tubing
and waste container must be chemical resistant and able
to withstand high temperatures (>150°C / 302°F).
When the bath has been installed it must be filled with an
appropriate liquid. When working with water the bath
should be filled to 1 cm below the lid. For oil the bath
should be filled to not more than 5 cm below the lid.
Depending on the operating temperature the liquid level
in the bath should be observed and excessive fluid should
be removed.
The liquid level should be maintained between 1 and 3
cm below the lid during normal operation.
A lower level than 5 cm below the lid may damage the
heaters. A high bath level can cause overflow and might
also damage the bath insulation.
H
2
O
FILLED 5 CM
BELOW THE
LID
MAX.
5 CM
MAX.
1 CM
Oil
Connect outlet
to waste
container
TUBING AND
CONTAINER:
CHEMICAL RESISTANT
WITHSTAND HIGH
TEMPERATURE