Van ’t Hoffstraat 12
2665 JL Bleiswijk, The Netherlands
T. 31 (0) 10 522 43 73
tv2000-4000mkI-4000DC.docx Rev. 1.06 UK 0718
E-mail: [email protected]
Website: www.tamson.com
Tamson Instruments bv
Page 16/33
ISO 9001 : 2015
NL/PRO 238239125
VAT: NL 80 66 34 984 B01
Bank account no.:
NL28 INGB 0007 350 370
NL95 RABO 0160100046
Chamber of commerce 27 16 95 41
6.5
Front panel layout
7
OPERATING THE BATH
When the bath is ready for use it can be switched on by
pressing the mains switch. The bath will initialise and
calibrate itself.
7.1
Overview menu items
-
Set point
-
Offset (press: <-5.00 .. +5.00°C resolution 0.01°C)
-
Max Power (press: low, med, hi, max)
-
Boost heater (press on / off)
-
Time const. (press: fast, medium slow, precise)
-
PID parameter :
PID set 1,
PID set 2,
PID set 3,
PID set 4, for use with RS232
Each PID set offers settings for
- Proportional band value
(Pb=1/P where P is proportional value)
- Integral value
- Differential value
-
Backlight (LCD)
-
Temp units
-
Baudrate
-
SP Offset
-
Restart
Item
Description
Function
1
Thermostat
Over-temperature protection
2
Display
Shows bath parameters
3
Switch
Menu
4
Switch
Down
5
Switch
Up
7
LED
Heater I
8
LED
Heater II
9
LED
Offset entered
10
LED
System - Error
11
Switch
Mains switch
12
Fuse
Protects stirrer motor
1
3
8
7
10
9
2
5
4
11
12