Van ’t Hoffstraat 12
2665 JL Bleiswijk, The Netherlands
T. 31 (0) 10 522 43 73
tv2000-4000mkI-4000DC.docx Rev. 1.06 UK 0718
E-mail: [email protected]
Website: www.tamson.com
Tamson Instruments bv
Page 21/33
ISO 9001 : 2015
NL/PRO 238239125
VAT: NL 80 66 34 984 B01
Bank account no.:
NL28 INGB 0007 350 370
NL95 RABO 0160100046
Chamber of commerce 27 16 95 41
Menu item "Offset 0.005"
This menu item offers an additional offset of 0.005°C. The
value of 0.005°C is added to the selected SP value. The
SP can be selected with 0.01°C accuracy.
Menu item "Restart"
Restarts system and activates tuning
8.1
PID settings
Default settings
P = 2 .. 25
I = 0 .. 25
D = 0
Following shows influence on stability with different PID
settings
Temperature
:
75°C
Proportional band (Pb) :
25
Integrator
:
16
Differentiator
:
0
Min/max
:
± 0.015
Temperature
:
75°C
Proportional band (Pb) :
50
Integrator
:
25
Differentiator
:
0
Min/max
:
± 0.018°C
Temperature
:
75°C
Proportional band (Pb) :
100
Integrator
:
25
Differentiator
:
0
Min/max
:
± 0.045°C
SP 75°C
0.029
± 0.015
SP 75°C
0.036
± 0.018
SP 75°C
0.089
± 0.045