Calibration
170
Output Voltage Control with External Resistance(B)
1. Disconnect the load wires and the remote sensing
wires. Connect the digital multimeter to the output
terminal.
Turn the POWER switch ON and OUTPUT OFF in
advance.
Power Supply
+
-
V
+
-
Digital
Multimeter
Calibration Circuit
4
17
External
Control Connector
External Resistance
0 to 10k
Ω
2. Set the external resistance to 0Ω and the constant
current setting value to [ yA ] and set the output to
ON.
30-200
30-400
60-100
60-200
[yA]
1.0A
1.0A
1.0A
1.0A
500-12
500-24
1000-6
1000-12
[yA]
0.10A
0.10A
0.010A
0.10A
On how to set, refer to Output Voltage Control with
External Resistance (B).(page139)
3. Press the FUNC. key.
4. Turn the VOLTAGE dial to display the setting item
number 28 in the Number Display (voltage).
* At that time, turn off the Number Display (current).
5. Adjust so that the output voltage becomes
"0.000V" with the CURRENT dial.
6. Press the FUNC. key to settle the offset calibration
value of output voltage control with external
resistance (B).
Содержание HX-S-G4 Series
Страница 157: ...Special Loads 157 This page is blank...
Страница 161: ...Maintenance 161 This page is blank...
Страница 197: ...Specifications 197 External Diagrams 1 HX S 030 200G4 F I...
Страница 198: ...Specifications 198 2 HX S 030 400G4 F I...
Страница 199: ...Specifications 199 3 HX S 060 100G4 F I...
Страница 200: ...Specifications 200 4 HX S 060 200G4 F I...
Страница 201: ...Specifications 201 5 HX S 0500 12G4 F I...
Страница 202: ...Specifications 202 6 HX S 0500 24G4 F I...
Страница 203: ...Specifications 203 7 HX S 01000 6G4 F I...
Страница 204: ...Specifications 204 8 HX S 01000 12G4 F I...