Calibration
185
7. Set the constant voltage setting value to [ xV ] and
the constant current setting value to [ yA ] and set
the output to ON.
30-200
30-400
60-100
60-200
[xV]
1.00V
1.00V
1.00V
1.00V
[yA]
200.0A
400.0A
100.0A
200.0A
500-12
500-24
1000-6
1000-12
[xV]
1.0V
1.0V
1V
1V
[yA]
12.00A
24.00A
6.000A
12.00A
See Operating as Constant Current Power Supply
under Basic Operations on how to set. (page29)
8. Press the FUNC. key.
9. Turn the VOLTAGE dial to display the setting item
number 27 in the Number Display (voltage).
* At that time, the current measured value is displayed
in the Number Display (current).
* The figure above is an example of connecting (HX030-200G4)
10. Adjust the CURRENT dial so that the current
measured value displayed in the Number Display
(current) is the same as the actual output current
value (digital multimeter measurement).
* The figure above is an example of connecting (HX030-200G4)
11. Press the FUNC. key to settle the full-scale
calibration value in the ammeter.
12. This completes offset/full-scale calibration of
ammeter.
Set the constant voltage setting value and the
constant current setting value to the original ones
and start using.
Содержание HX-S-G4 Series
Страница 157: ...Special Loads 157 This page is blank...
Страница 161: ...Maintenance 161 This page is blank...
Страница 197: ...Specifications 197 External Diagrams 1 HX S 030 200G4 F I...
Страница 198: ...Specifications 198 2 HX S 030 400G4 F I...
Страница 199: ...Specifications 199 3 HX S 060 100G4 F I...
Страница 200: ...Specifications 200 4 HX S 060 200G4 F I...
Страница 201: ...Specifications 201 5 HX S 0500 12G4 F I...
Страница 202: ...Specifications 202 6 HX S 0500 24G4 F I...
Страница 203: ...Specifications 203 7 HX S 01000 6G4 F I...
Страница 204: ...Specifications 204 8 HX S 01000 12G4 F I...