Using in Series/Parallel Operation
126
8. Press the FUNC. key of the Slave.
9. Turn the VOLTAGE dial of the Slave unit to
display the setting item number 38 in the Number
Display (voltage).
* At that time, turn off the Number Display (current).
10. Adjust the CURRENT dial of the Slave so that the
display of the Slave unit’s digital multimeter for
output voltage measurement is the same as the
display of the Master unit’s digital multimeter for
output voltage measurement.
11. Press the FUNC. key of the Slave unit to set the
full-scale calibration value of output voltage
setting.
12. This completes offset full-scale calibration for the
output voltage of the Slave unit in series
operation.
At the Master unit, set the constant voltage
setting value and the constant current setting
value to the original ones and start using.